Assessment of precision and accuracy of bone mineral measurement by using spectral CT imaging:phantom study

( views:441, downloads:396 )
Author:
No author available
Journal Title:
Chinese Journal of Radiology
Issue:
11
DOI:
10.3760/cma.j.issn.1005-1201.2014.11.010
Key Word:
骨密度;体层摄影术,X线计算机;对比研究;Bone mineral density;Tomography,X-ray computed;Comparative study

Abstract: Objective To assess the precision and accuracy of bone mineral measurement utilizing spectral CT imaging with a European Spine Phantom (ESP).Methods Within the ESP named Num.145,there are three orderly vertebral bodies which were marked by L1 with Hydroxyapatite(HAP)of 50.0 mg/cm3,L2 with HAP of 100.0 mg/cm3,and L3 with HAP of 200.0 mg/cm3.The ESP was scanned by using gemstone spectral imaging (GSI) model ten times.Using GSI viewer,measurements of HAP-water concentration and HAP-fat concentration were done.The precision and accuracy of each parameter were evaluated by the results of relative standard deviation (RSD) and relative error (RE),respectively.The difference between each parameter with the real value was evaluated statistically by one-sample t test.The differences among the three vertebral bodies' same parameter were evaluated statistically by paired t test.Results The results of HAP-water concentration in L1,L2,L3 were (54.5 ±4.2),(113.3 ± 2.7) and (219.7 ± 7.7) mg/cm3,each of which was significantly higher than the real value(t values were 3.400,15.874 and 8.155,P<0.01).And,the RSD values were 7.67%,2.34% and 3.48%;the RE values were 5.3%,12.9% and 9.8%.The results of HAP-fat concentration in L1,L2,L3 were (91.9±4.0),(149.5±2.5) and (254.2±7.2) mg/cm3,each of which was significantly higher than the real value(t values were 33.489,62.399 and 23.578,P<0.01).And the RSD values were 4.31%,1.68% and 2.86%; the RE values were 81.0%,49.1% and 27.0%.In every vertebral body,the result of HAP-water concentration was significantly lower than the result of HAP-fat concentration (P<0.01).Conclusions By utilizing spectral CT imaging,the precision and accuracy of bone mineral measurement were poor.Therefore,this method is not applicable for clinical measurement.

WanfangData CO.,Ltd All Rights Reserved
About WanfangData | Contact US
Healthcare Department, Fuxing Road NO.15, Haidian District Beijing, 100038 P.R.China
Tel:+86-010-58882616 Fax:+86-010-58882615 Email:yiyao@wanfangdata.com.cn