Abstract: @@ 一、故障现象开机即出现报警,故障信息显示"MIN kV"(千伏过低),摄片无法进行.二、故障分析与检修首先分析千伏预置电路原理.如随机电路原理图05275及图1所示,通过操作控制面板B4上kV(+)、kV(-)按键,千伏预置信号经输入接口电路IC2输入微处理器(MCPU),MCPU经输出接口电路IC5、光耦OC7、OC8在输出端output 6、output 7分别输出kV+ selection、kV-selection指令信号(参见随机原理图05274),此信号印刷电路板B2的运放IC2、IC3(参见随机原理图05272).