Comparison of prospective electrocardiography-gating high-pitch mode and without electrocardiography-synchronization high-pitch mode acquisition for the image quality and radiation doses of the aortic using dual-source CT

( views:0, downloads:0 )
Author:
No author available
Journal Title:
Chinese Journal of Radiology
Issue:
4
DOI:
10.3760/cma.j.issn.1005-1201.2013.04.003
Key Word:
主动脉造影术;辐射剂量;体层摄影术,X线计算机;质量控制;Aortography;Radiation dose;Tomography,X-ray computed;Quality control

Abstract: Objective To evaluate the application of prospective ECG-gating Flash spiral scan mode dual-source CT in aortography,and compare it's image quality and radiation dose with without ECG-synchronization high-pitch spiral scanning mode.Methods Fifty consecutive patients (Group A) with suspected aortic dissection or after operations for the aortic dissection were scanned with prospective ECG-gated high-pitch scan and another 50 consecutive patients(Group B) were analyzed by non-ECG-gated high-pitch scan.Image quality of the aortic was assessed by two independent readers.Image noise was measured,radiation dose estimates were calculated.The imaging quality of the aortic and the radiation dose were compared with Mann-whitney U and t test.Results The average image quality score [(1.18 ± 0.40) in group A and (1.23 ±0.31) in group B] showed no significant difference between group A and group B(U =1.20,P =0.23).The mean radiation dose of group A was lower than that of group B [(1.49 ± 0.38) mSv in group A,(2.79 ± 0.54) mSv in group B,t =13.677,P < 0.05].Conclusion Prospective ECG-gated dualsource CT Flash spiral scanning with low radiation dose and good image quality in the aortic dissection with high value of clinical application.

WanfangData CO.,Ltd All Rights Reserved
About WanfangData | Contact US
Healthcare Department, Fuxing Road NO.15, Haidian District Beijing, 100038 P.R.China
Tel:+86-010-58882616 Fax:+86-010-58882615 Email:yiyao@wanfangdata.com.cn