島津3000 TX型CT机复合故障一例

( views:391, downloads:0 )
Author:
No author available
Journal Title:
CHINESE JOURNAL OF RADIOLOGY
Issue:
5
DOI:
10.3760/j.issn:1005-1201.2001.05.023
Key Word:
No keyword available

Abstract: @@故障现象:机器起动正常,按下扫描开始键,当操作台上X-ready指示灯一亮,在扫描室可听到一声巨响,随之X-ready指示灯熄灭,机器停止工作.故障分析检修:检查X线控制器,发现控制驱动电路板上指示灯均不亮,电源继电器(MSW1)掉闸.选择X线控制器方式(local),闭合电源开关,控制驱动电路板上指示灯正常.按下ready开关,进行X线准备,X-ready指示灯一亮,又是一声巨响,再次发生电源中断.马上利用充电电压指示表(CH.kV)检查阴阳两极高压充电电容电压,阴极侧为145 kV,正常;阳极侧为45kV,异常.初步判定阳极侧高压充电电容击穿,高压放电.把高压发生器吊出油面,用摇表对阳极侧高压充电电容检查,见该电容击穿.更换同规格电容,按ready开关,X-ready指示灯亮,无异常声音,按X-ray开关,进行X线曝光,有X线产生.