Abstract: 患者女,68岁,因快-慢综合征于2010年2月12日植入双腔起搏器(Zephyr 5826,美国圣犹达公司),参数如下:起搏模式DDD,下限频率50 次/min,休息频率45 次/min,心室后心房不应期(post ventricular atrial refractory period,PVARP) 250 ms,心室后心房空白期(post ventricular atrial blanking period,PVAB)120 ms,起搏的房室间期/感知的房室间期(paced atrioventricular interval/sensed atrioventricular interval,PAV/SAV)350/325 ms,最大跟踪频率100次/min,最大传感器频率105次/min,心房颤动抑制( AF suppression)功能打开,超速起搏周期15,最大超速起搏频率105次/min.患者心悸时记录到心电图,为心房扑动伴间断心房起搏,不除外心房、心室感知功能不良;起搏器程控时腔内电图,结合起搏器腔内电图特点考虑图1并非起搏器心房、心室感知功能不良,而是起搏器的AF suppression功能在发挥作用.