Initial application of prospective electrocardiography-triggering high-pitch spiral acquisition by dualsource CT in children with congenital heart diseases

( views:447, downloads:0 )
Author:
JI Xiao-peng(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
CHEN Li-guang(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
WANG Xi-ming(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
CHENG Zhao-ping(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
NIE Pei(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
QIAO Hong-yan(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
LI Hai-ou(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
YANG Shi-feng(Department of CT, Shandong Medical Imaging Research Institute, Shandong University School of Medicine, Jinan 250012, China)
Journal Title:
Chinese Journal of Radiology
Issue:
Volume 46, Issue 10, 2012
DOI:
10.3760/cma.j.issn.1005-1201.2012.10.012
Key Word:
Heart defects,congenital; Tomography, spiral computed; Radiation dosage

Abstract: Objective To explore the clinical value of prospective electrocardiography-triggering high-pitch spiral acquisition by dual-source CT (HP mode ) in children with congenital heart diseases (CHD).Methods Thirty-six patients ( mean age:0.67 years,range:1 month to 2 years and 6 months; 25 males;mean weight:7.7 kg)underwent prospective ECG-triggering high-pitch spiral acquisition.The dose length product (DLP) was recorded to calculate effective dose ( ED ).Two experienced radiologists made diagnosis and assessed the overall image quality by a five-point scale independently.The consistency of their judgment was determined by Kappa statistics.Surgeries were performed in 29 patients,and conventional cardiac angiography(CCA) was performed in 7 patients.The accuracy,sensitivity and specificity of HP mode were calculated based on the surgical or CCA findings.Results The DLP was (5.12 ± 1.64) mGy.cm with a range of 2.4 mGy · cm to 9.2 mGy · cm and ED was (0.125 ± 0.027 ) mSv ranging from 0.078 mSv to 0.179 mSv.The average subjective image quality score was 4.2 ± 0.7.All images were diagnosable.Two radiologists made good agreement ( K =0.774,P < 0.05 ).The accuracy,sensitivity and specificity of HP mode were 98.9%,94.2% and 99.8%,respectively.Conclusion Low radiation dose with the accurate diagnosis of anomalies makes HP mode a new choice for children with CHD.

  • [1]Lell M, Marwan M,Schepis T,et al. Prospectively ECGtriggered high-pitch spiral acquisition for coronary CT angiographyusing dual source CT: techniqueand initial experience.Eur Radiol,2009,19:2576-2583.
  • [2]Achenbach S,Marwan M,Ropcrs D,et al.Coronary computed tomography angiography with a consistent dose below ImSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition.Eur Heart J,2010,31:340-346.
  • [3]Achenbach S,Goroll T,Seltmann M,et al.Detection of coronary artery stenoses by low-dose prospectively ECG-triggered high-pitch spiral coronary CT angiography.JACC Cardiovasc Imaging,2011,4:328-337.
  • [4]王怡宁,李烁,孔令燕,等.低剂量双源CT前瞻性心电触发大螺距扫描冠状动脉造影的图像质量和放射剂量.中国医学科学院学报,2010,32:597-600.
  • [5]Hall EJ.Lessons we have learned from our children:cancer risks from diagnostic radiology.Pediatr Radiol,2002,32:700-706.
  • [6]Thomas KE,Wang B.Age-specific effective doses for pediatric MSCT examinations at a large children's hospital using DLP conversion coefficients:a simple estimation methocd.Pediatr Radiol,2008,38:645-656.
  • [7]Winklehner A,Karlo C,Puippe G,et al. Raw data-based iterative reconstruction in body CTA:evaluation of radiation dose saving potential.Eur Radiol,2011,21:2521-2526.
  • [8]Tsai IC, Lee T,Chen MC,et al. Visualization of neonatal coronary arteries on multidetector row CT:ECG-gated versus nonECG-gated technique.Pediatr Radiol,2007,37:818-825.
  • [9]Achenbach S,Marwan M,Schepis T,et al.High-pitch spiral acquisition:a new scan mode forcoronary CT angiography. J Cardiovasc Comput Tomogr,2009,3:117-121.
  • [10]Jin KN,Park EA,Shin CI,et al.Retrospective versus prospective ECG-gated dual-source CT in pediatric patients with congenital heart diseases:comparison of image quality and radiation dose.Int J Cardiovasc Imaging,2010,26:63-73.
  • [11]Paul JF,Rohnean A,Elfassy E,et al.Radiation dose for thoracic and coronary step-and-shoot CT using a 128-slice dual-source machine in infants and small children with congenital heart disease.Pediatr Radiol,2011,41:244-249.
  • [12]Lee T,Tsai IC,Fu YC,et al. Using muhidetector-row CT in neonates with complex congenital heart disease to replace diagnostic cardiac catheterization for anatomical investigation:initial experiences in technical and clinical feasibility.Pediatr Radiol,2006,36:1273-1282.